Cath Copilot Book a demo

Guide · checked November 2026

Cath lab audit checklist: what to check and against which standard

A cath lab audit checklist should test the lab against the standards that actually apply: IR(ME)R employer's procedures, entitlement and training records, NatSSIPs safety checks, infection prevention and control, complete records and national audit data, and traceable devices and consumables. The evidence should come from real cases traced end to end, not only from policies on file.

Cath Copilot audit log viewed by the quality lead, showing a verified hash chain of actions
Audit log — append-only and hash-chained. Sign-ins, record queries, AI drafts, and denials are all recorded, and the quality lead can verify the chain.

Radiation: IR(ME)R

  • Employer's procedures (IR(ME)R Schedule 2) that describe what the cath lab actually does, not only what the trust's radiology department does.
  • Entitlement: who may act as referrer, practitioner and operator, and for which procedures.
  • Training records for every practitioner and operator. Regulation 17(4) requires an up-to-date record, available for inspection, showing when adequate training was completed and its nature.
  • Referral guidelines made available to referrers.
  • Patient dose recorded for each procedure and reviewed against local diagnostic reference levels.
  • Equipment quality assurance, and a plan for replacing ageing equipment.
  • Significant accidental or unintended exposures investigated and notified as required.

CQC enforces IR(ME)R in England. In its 2021/22 annual report, describing a neurointerventional inspection programme whose themes it said may be relevant to other cardiology or radiology interventional services, CQC regularly found employer's procedures that were not representative of the practice in the department, often because they were written generically for several services. One example was a patient identification procedure with no reference to the WHO checklist. It also reported that referral guidelines had often not been implemented, adding that it had seen this in cardiology inspections too, and it flagged ageing equipment as a risk.

Safety checks: NatSSIPs

  • A LocSSIP for each procedure type, written from NatSSIPs 2.
  • Sign in, time out and sign out done with the whole team present, observed rather than only ticked.
  • Implant verification before a device is opened.
  • Debriefs held and their actions followed up.

Infection prevention and control

NHS England says its National Infection Prevention and Control Manual should be adopted as mandatory guidance in NHS settings and wherever NHS services are delivered. Audit the lab's practice, such as hand hygiene, aseptic technique, cleaning between cases and waste handling, against the NIPCM and your local policy.

Records and national audit

  • Procedure note complete and signed, with contrast volume, dose, devices and complications stated.
  • Complications recorded explicitly, including when there were none.
  • National audit records (for example NAPCI) complete, consistent with the note, and submitted on time.
  • Handover documented and acknowledged by the receiving team.

Devices and consumables

  • Every implant traceable to the patient by make, model and lot.
  • Stock in date, with expired items removed.
  • Used devices match between the procedure record, the inventory scan and the audit return.

Make the audit useful

Pick a sample of recent cases and trace each one from referral to discharge, checking the documents and the records against each other. Record the gaps, agree owners and dates, and re-audit to show the loop has closed.

Common traps

  • A policy on file is not evidence of practice: CQC has found procedures that did not match what departments did.
  • A ticked checklist is not the same as a time out done properly.
  • A blank complications field reads as "none" when nobody stated it.

Where Cath Copilot fits

Cath Copilot leaves an evidence trail for the record-based parts of an audit. Every action lands in an append-only, hash-chained audit log. Reconciliation overrides need a reason, and the quality review queue lists missing registry fields, potential events for review and reconciliation overrides. The audit return shows the source of every field and only the quality lead can export it. Pre-case flags record who resolved them and why.

It does not hold IR(ME)R employer's procedures, entitlement or training records, and it does not audit infection control practice. Cath Copilot is a prototype on synthetic data only. It is not a medical device and is not for clinical use.

Cath Copilot overview screen Cath Copilot: An AI copilot for the cath labWatch the demo on synthetic data, then see how it works →

Questions

Who inspects IR(ME)R compliance in England?

CQC enforces IR(ME)R in England, in both NHS and independent services.

What training records does IR(ME)R require?

Regulation 17(4) requires the employer to keep an up-to-date record of relevant training for all practitioners and operators, showing the dates and nature of the training, available for inspection.

Should a cath lab audit use real cases?

Yes. Tracing real cases shows whether practice matches the written procedures, which is a gap CQC reports finding in interventional services.

Sources

Official pages, checked November 2026. Guidance changes, so check the current version before you rely on it.

General information for healthcare managers and clinical teams, checked November 2026. It is not legal, regulatory or clinical advice; follow the current official guidance and your organisation's own policies. No patient data is used in these examples. Cath Copilot is a prototype running on synthetic data only. It is not a medical device and is not for clinical use. It recommends, drafts, and routes; it never diagnoses, orders, or decides. A clinical safety and information-governance process precedes any use with real patient data.

See Cath Copilot on synthetic data

Cath Copilot is a pilot. Book a demo and we'll walk through it on synthetic data, then talk about what a pilot under your own governance would need.

Book a demo

Or email hi@zapxlabs.com.